The following guidelines are recommended for
the setup.
- For CE, an external, 5V power supply is
used to provide power to the device under test (DUT) through the line impedance
stabilization networks (LISNs). As required by the CISPR 32 test standard, the LISNs are
placed and connected to a metal conductive floor and must be at least 80cm away from the
DUT. The DUT must then be placed on a non-conductive table at least 10cm away from the
external power supply.
- For RE, a low noise power supply is used
to power the DUT. It is very important that the connection from the power supply to the
DUT is short and tightly coupled. As seen in Figure 2-3, a twisted pair is used to connect both.